Description
Model: AG-Staray 5000
Brand: Advanced Health Canada
Country of origin:- Canada
Medical Digital Radiography System – Double Column DR System AG-Staray 5000
specification
| 1 | Flat panel detector |
| 1.1 | Amorphous silicon wired flat panel detector |
| 1.2 | Effective pixels: 9,<> |
| 1.3 | Pixel matrix: 3071*3071 pixels |
| 1.4 | A/D conversion: 16 bits |
| 1.5 | Effective Area: 17″×17″ |
| 1.6 | Spatial resolution: 3.7 l/mm |
| 1.7 | Imaging time: 6.82s |
| 2 | Mechanical structure: intelligent electric double column |
| 2.1 | Bracket movement mode: electric |
| 2.2 | Vertical range (tube assembly): 1400mm (range: 400mm~1800mm) |
| 2.3 | Horizontal range: 1200mm |
| 2.4 | Tube assembly rotation range: 360° |
| 2.5 | Vertical range (detector assembly): 1400mm (range: 400mm~1800mm) |
| 2.6 | Intelligent collision avoidance system (intelligent protection) |
| 2.7 | Intelligent Automatic Height Recognition System (Smart POS) |
| 3 | High voltage generators |
| 3.1 | Maximum operating frequency: 200kHz |
| 3.2 | Power: 50kW |
| 3.3 | kV range: 40kV -150kV |
| 3.4 | mA range: 10mA -630mA |
| 3.5 | mAs range: 0.1mA -630mA |
| 3.6 | High-voltage generator intelligent network voltage real-time monitoring system (intelligent monitoring) |
| 4 | X-ray tube |
| 4.1 | Focus size: 0.6mm/1.2mm |
| 4.2 | Focus power: Small focal length: 20kW Large focal length: 50kW |
| 4.3 | Maximum heat capacity: 230kHU |
| 4.4 | Maximum peak voltage: 150kV |
| 4.5 | Anode speed: 2700 rpm |
| 4.6 | Anode heat dissipation rate: 667HU/s |
| 5 | Filter grid |
| 5.1 | Ratio: 10:1 |
| 5.2 | Focus distance: 100 cm |
| 5.3 | Size: 18inch * 18inch |
| 5.4 | Fixed type |
| 6 | LED Collimator |
| 6.1 | The time limit of the indicator: 30±3s |
| 6.2 | Radiation field indicator: 150W, 24VAC |
| 6.3 | Inherent filtration: 1.0 mm Al. EQ. |
| 7 | Table |
| 7.1 | Photoelectric controlled four-way suspended table |
| 7.2 | Length ×wide of table surface:2200*800mm |
| 7.3 | Movement stroke (longitudinal):840mm |
| 7.4 | Movement stroke (horizontal):240mm |
| 7.5 | Movement range of detector(horizontal):400mm |
| 7.6 | Bed height:700mm |
| 7.7 | Load bearing:≥200kg |
| 7.8 | Bed-detector distance:80mm |
| 8 | Workstation |
| 8.1 | DELL 5050MT,Intel Pentium G4560 processor, 4G/500G/DVDRW/WIN10,E2216H 21.5 inch high contrast display |
| 8.2 | Software function(DR workstation software can control the stand) Intelligent control system: You can control the operation of the stand freely through computer software, and control the movement and rotation of the tube. |
| 8.3 | Software function(software direct exposure control) |
| 8.3.1 | Support customization for dose template; |
| 8.3.2 | The selection part and position template, automatically bring out the matching dose; it can also provide early warning when voltage fluctuation is large; Selection shape of the patient, automatically select the matching dose, and adjust to the dose online; |
| 8.3.3 | Control the size of focus; |
| 8.3.4 | Display generator’s status transparent; |
| 8.3.5 | Refusal and acceptance of images; |
| 8.3.6 | Support automatic sending of images. |
| 8.3.7 | AEC: optional |
| 8.3.8 | Image stitching: optional |
| 8.4 | Software function(patient management) |
| 8.4.1 | Remote direct access to patient information in RIS with worklist protocol; Manually create the patient examination information; |
| 8.4.2 | Emergency quick registration; |